Comparative study between EDXRF and ASTM E572 methods using two-way ANOVA
Visualizar/abrir
Data
2018Tipo
Abstract
Comparison with reference method is one of the necessary requirements for the validation of non-standard methods. This comparison was made using the experiment planning technique with two-way ANOVA. In ANOVA, the results obtained using the EDXRF method, to be validated, were compared with the results obtained using the ASTM E572-13 standard test method. Fisher's tests (F -test) were used to comparative study between of the elements: molybdenum, niobium, copper, nickel, manganese, chromium and v ...
Comparison with reference method is one of the necessary requirements for the validation of non-standard methods. This comparison was made using the experiment planning technique with two-way ANOVA. In ANOVA, the results obtained using the EDXRF method, to be validated, were compared with the results obtained using the ASTM E572-13 standard test method. Fisher's tests (F -test) were used to comparative study between of the elements: molybdenum, niobium, copper, nickel, manganese, chromium and vanadium. All F -tests of the elements indicate that the null hypothesis (Ho) has not been rejected. As a result, there is no significant difference between the methods compared. Therefore, according to this study, it is concluded that the EDXRF method was approved in this method comparison requirement. ...
Contido em
Journal of physics : conference series. Bristol. Vol. 975 (2018), 012004
Origem
Estrangeiro
Coleções
-
Artigos de Periódicos (40304)Engenharias (2437)
Este item está licenciado na Creative Commons License