Stability-indicating method for determination of tizanidine hydrochloride by LC-CAD method using chemometric approach
Visualizar/abrir
Data
2017Autor
Tipo
Assunto
Abstract
A new LC method for tizanidine hydrochloride in tablet dosage form using a charged aerosol detector (CAD) is described. The influence of various parameters on chromatographic system was investigated by factorial designs and Derringer's desirability. Chromatographic conditions were: mobile phase constituted of acetonitrile and ammonium acetate buffer 17 mM (60:40; v/v), column oven at 39 °C and flow rate 0.8 mL.min-1 performed on Acclaim Trinity P1 column UV at 230 nm. Thus, it was possible to v ...
A new LC method for tizanidine hydrochloride in tablet dosage form using a charged aerosol detector (CAD) is described. The influence of various parameters on chromatographic system was investigated by factorial designs and Derringer's desirability. Chromatographic conditions were: mobile phase constituted of acetonitrile and ammonium acetate buffer 17 mM (60:40; v/v), column oven at 39 °C and flow rate 0.8 mL.min-1 performed on Acclaim Trinity P1 column UV at 230 nm. Thus, it was possible to validate a simple method to assay the tizanidine and its counter-ion in three formulations (drug reference, generic and manipulated). The present method showed specificity when challenged by forced degradation and excipients. Finally, the method was compared with USP monograph method demonstrating equivalence in assay evaluation. ...
Contido em
Drug analytical research. Porto Alegre (RS). Vol. 1, n. 2 (2017), p. 61-66
Origem
Nacional
Coleções
-
Artigos de Periódicos (46876)Ciências da Saúde (12810)
Este item está licenciado na Creative Commons License