Mostrar registro simples

dc.contributor.authorMurliky, Lucaspt_BR
dc.contributor.authorOliveira, Gustavo dept_BR
dc.contributor.authorSousa, Fernando Rangel dept_BR
dc.contributor.authorBrusamarello, Valner Joaopt_BR
dc.date.accessioned2022-10-27T04:51:19Zpt_BR
dc.date.issued2022pt_BR
dc.identifier.issn1424-8220pt_BR
dc.identifier.urihttp://hdl.handle.net/10183/250459pt_BR
dc.description.abstractThis work presents an inductive wireless power transfer system for powering an endoscopy capsule supplying energy to power electronic devices allocated inside a capsule of ≈26.1 mm × 9 mm. A receiver with three coils in quadrature with dimensions of ≈9 mm × 9 mm × 10 mm is located inside the capsule, moving freely inside a transmitter coil with 380 mm diameter through translations and revolutions. The proposed system tracks the variations of the equivalent magnetic coupling coefficient compensating misalignments between the transmitter and receiver coils. The power on the load is estimated and optimized from the transmitter, and the tracking control is performed by actuating on a capacitance in the matching network and on the voltage source frequency. The proposed system can prevent load overheating by limiting the power via adjusting of the magnitude of voltage source VS. Experimental results with uncertainties analysis reveal that, even at low magnetic coupling coefficients k ranging from (1.7 × 10−3 , 3.5 × 10−3 ), the power on the load can be held within the range of 100–130 mW. These results are achieved with any position of the capsule in the space, limited by the diameter of the transmitter coil and height of 200 mm when adjusting the series capacitance of the transmitter in the range (17.4, 19.4) pF and the frequency of the power source in the range (802.1, 809.5) kHz.en
dc.format.mimetypeapplication/pdfpt_BR
dc.language.isoengpt_BR
dc.relation.ispartofSensors. Basel. Vol. 22, n. 18 (Sept. 2022), art. 6924, 17 p.pt_BR
dc.rightsOpen Accessen
dc.subjectEndoscopic capsuleen
dc.subjectTransferência de energia sem fiopt_BR
dc.subjectWireless power transferen
dc.subjectWPT dynamic trackingen
dc.titleTracking and dynamic tuning of a wireless powered endoscopic capsulept_BR
dc.typeArtigo de periódicopt_BR
dc.identifier.nrb001151867pt_BR
dc.type.originEstrangeiropt_BR