Mostrar registro simples

dc.contributor.authorCasco, Márcia Flores dept_BR
dc.contributor.authorMoraes, Maria Antonieta P.pt_BR
dc.contributor.authorSouza, Emiliane Nogueira dept_BR
dc.contributor.authorCasco Raudales, Jose Jesuspt_BR
dc.contributor.authorSilva, Eneida Rejane Rabelo dapt_BR
dc.date.accessioned2014-01-29T01:53:59Zpt_BR
dc.date.issued2009pt_BR
dc.identifier.issn0104-1843pt_BR
dc.identifier.urihttp://hdl.handle.net/10183/86971pt_BR
dc.description.abstractA prática de reprocessamento, de reesterilização e de reutilização de artigos odonto-médico-hospitalares é ampla no País e no mundo, e tem como objetivo fundamental reduzir os custos dos procedimentos. Os dispositivos de insuflação de cateteres de angioplastia não fogem a essa regra de reutilização. A avaliação das características físicas e funcionais desses dispositivos após processamento para sua reutilização permanece pouco explorada na literatura. Objetivos: Avaliar se as características físicas e funcionais dos dispositivos de insuflação de cateteres de angioplastia, após os processos para a reutilização, são mantidas, bem como determinar a vida útil de tais dispositivos. Método: Estudo longitudinal contemporâneo. Incluíram-se dispositivos de insuflação de cateteres de angioplastia novos, utilizados no período de novembro de 2007 a março de 2008. Estudo realizado em hospital universitário, localizado em Porto Alegre, Rio Grande do Sul, Brasil. As avaliações foram realizadas em dois momentos: grupo A, dispositivos avaliados imediatamente após o primeiro uso; e grupo B, dispositivos avaliados após usos subsequentes. Resultados: Dos 36 dispositivos estudados, 11 (30,6%) foram desprezados logo após o primeiro uso por apresentarem problemas mecânicos e funcionais graves. Dos dispositivos restantes, apenas 8 (32%) no grupo A e 6 (24%) no grupo B mantiveram sua funcionalidade integralmente preservada. Esses instrumentais apresentaram alterações significativas, ainda que de pequena magnitude, nas aferições a partir de 12 atm. A média de reutilização foi de 1,7 ± 1,2 vez. Conclusões: Demonstrou-se, neste estudo, que cerca de 30% dos dispositivos de insuflação de cateteres de angioplastia se apresentaram inadequados após a primeira utilização. E somente um quarto dos remanescentes manteve suas características funcionais íntegras até o final de sua vida útil.pt_BR
dc.description.abstractBackground: Reuse of dental and medical devices is a worldwide common practice, and our country is not an exception; its aim is to reduce procedural costs. PTCA insufflation devices (or manometers) are frequently resterilized for clinical reuse purposes. Evaluations of their physical and mechanical properties are not described in current literature. Objective: Determine if the physical and functional characteristics of insufflation devices are maintained after reuse processing, as well as establish their life span. Methods: This was a prospective longitudinal study, carried out at university hospital in Porto Alegre, RS, Brazil. From November 2007 to March 2008, all new insufflation devices were included and separated in two different groups: Group A devices were evaluated immediately after their first use and Group B devices were evaluated after subsequent uses. Results: Of the 36 devices, 11 (30.6%) were discarded after the first use due to severe mechanical or functional problems. Of the remaining devices, only 8 (32%) in Group A and 6 (24%) in Group B kept totally preserved functionality. These devices presented significant changes, even though they were mild, in measurements after 12 atm. Mean reuse rate was 1.7 ± 1.2 times. Conclusions: Results showed that about 30% of PTCA insufflation devices were not in working conditions after the first use. Only a quarter of the remaining devices maintained intact functional characteristics to the end of their life span.en
dc.format.mimetypeapplication/pdfpt_BR
dc.language.isoporpt_BR
dc.relation.ispartofRevista brasileira de cardiologia invasiva. São Paulo. Vol. 17, n. 2 (abr./jun. 2009), p. 227-233pt_BR
dc.rightsOpen Accessen
dc.subjectPercutaneous transluminal coronary angioplastyen
dc.subjectAngioplastia coronária com balãopt_BR
dc.subjectEquipamentos e provisões hospitalarespt_BR
dc.subjectEquipment and hospital suppliesen
dc.subjectEquipment reuseen
dc.subjectReutilização de equipamentopt_BR
dc.titleAvaliação das propriedades físicas e funcionais dos dispositivos de insuflação de cateteres de angioplastia submetidos a processo para reutilizaçãopt_BR
dc.title.alternativeEvaluation of the physical and functional properties of angioplasty insufflation devices undergoing reuse processing en
dc.typeArtigo de periódicopt_BR
dc.identifier.nrb000708547pt_BR
dc.type.originNacionalpt_BR


Thumbnail
   

Este item está licenciado na Creative Commons License