Mostrar registro simples

dc.contributor.authorBarros, Sérgio Estelita Cavalcantept_BR
dc.contributor.authorJanson, Guilhermept_BR
dc.contributor.authorChiqueto, Kelly Fernanda Galvãopt_BR
dc.contributor.authorFerreira, Eduardo Silveirapt_BR
dc.date.accessioned2015-05-01T01:58:16Zpt_BR
dc.date.issued2014pt_BR
dc.identifier.issn1687-8728pt_BR
dc.identifier.urihttp://hdl.handle.net/10183/115745pt_BR
dc.description.abstractPurpose.This study evaluated the influence of recycling process on the torsional strength of mini-implants. Materials and Methods. Two hundred mini-implants were divided into 4 groups with 50 screws equally distributed in five diameters (1.3 to 1.7 mm): control group (CG): unused mini-implants, G1: mini-implants inserted in pig iliac bone and removed, G2: same protocol of group 1 followed by sonication for cleaning and autoclave sterilization, and G3: same insertion protocol of group 1 followed by sonication for cleaning before and after sandblasting (Al2O3-90 𝜇) and autoclave sterilization. G2 and G3 mini-implants were weighed after recycling process to evaluate weight loss (W). All the screws were broken to determine the fracture torque (FT). The influence of recycling process on FT and W was evaluated by ANOVA, Mann-Whitney, and multiple linear regression analysis. Results. FT was not influenced by recycling protocols even when sandblasting was added. Sandblasting caused weight loss due to abrasive mechanical stripping of screw surface. Screw diameter was the only variable that affected FT. Conclusions. Torsional strengths of screws that underwent the recycling protocols were not changed.Thus, screw diameter choice can be a more critical step to avoid screw fracture than recycling decision.en
dc.format.mimetypeapplication/pdfpt_BR
dc.language.isoengpt_BR
dc.relation.ispartofInternational Journal of Dentistry. Cairo. Vol. 2014 (2014), ID 424923, 5 p.pt_BR
dc.rightsOpen Accessen
dc.subjectImplantodontiapt_BR
dc.subjectOdontologiapt_BR
dc.titleEffect of recycling protocol on mechanical strength of used mini-implantspt_BR
dc.typeArtigo de periódicopt_BR
dc.identifier.nrb000949953pt_BR
dc.type.originEstrangeiropt_BR


Thumbnail
   

Este item está licenciado na Creative Commons License

Mostrar registro simples